Read RTD Sensors

Choose this function to measure temperatures with an RTD
probe or sensor.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select RTD for the FUNCTION, °F or °C for the UNITS
and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper
10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850
is the most common RTD type.
You must also select the WIRE
MODE for 2W, 3W or 4W to match the 2, 3 or 4 wires on the RTD
sensor.

Disconnect all wires from the sensor and connect the PIE
820-ELITE to the inputs of the device using 2, 3 or 4 wires.

Signals above the maximum scale are limited by protection
circuitry with “OVERRANGE” flashed on the display and the red
OVERLOAD LED lit.

The PIE 820-ELITE measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch

q

to MAX to see the highest reading and to

MIN to see the lowest reading. Press and hold the knob

e

to

clear the MAX and MIN readings.