Typical error conditions, Page 17, Ma pwr meas – Practical Instrument Electronics 820-ELITE User Manual

Page 19: Using ground leak detection, Process indicator

Advertising
background image

Page 17

The PIE 820-ELITE is
supplying the loop voltage.
There is a control loop
error. This may be a
transmitter (set for upscale
burnout) with a bad or
missing sensor, or a short
in the loop.

The PIE 820-ELITE is
supplying the loop voltage.
A calibrated transmitter is
limiting the loop current to
12.00 mA. An additional
0.28 mA is not controlled
by the transmitter and is
leaking somewhere in the
loop.

LEAK ERROR =
OVER RANGE
PWR MEAS

LEAK ERROR = 00.28

12.280

mA

PWR MEAS

Typical Error Conditions