Specifications – Practical Instrument Electronics 820-ELITE User Manual

Page 35

Advertising
background image

Page 33

Read mA
Ranges and Resolution 0.000 to 24.000 mA or -25.00 to 125.00%

of 4-20 mA

Accuracy

≤ ± (0.02 % of Reading + 0.003 mA)

Voltage burden

≤ 2V at 24 mA

Overload/Current

limit protection

25 mA nominal

Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution

0.000 to 24.000 mA or -25.00 to 125.00%

of 4-20 mA

Accuracy

≤ ± (0.02 % of Reading + 0.003 mA)

Loop compliance voltage ≥ 24 DCV @ 20.00mA
Loop drive capability

1200 Ω at 20 mA for 15 hours nominal;
950 Ω with Hart Resistor enabled

mA 2-Wire Transmitter Simulation
Accuracy

Same as Source/Power & Measure

Voltage burden

≤ 2V at 20 mA

Overload/Current limit

protection

24 mA nominal

Loop voltage limits

2 to 60 VDC (fuse-less protected from

reverse polarity connections)

Specifications

DC Voltage Read
Range and Resolution

±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00

to 60.00 V DC

Accuracy

≤ ± (0.02 % of Reading + 0.01% Full Scale)

Input resistance

≥ 1 MΩ