Page 19

Process Calibrator

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

READ

mV

8.203

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

Voltage Output Signal

Controller

Transmitter

Power Supply

Read mV/V

Choose this function to measure from -99.999 to 99.999
millivolts, -99.999 to +99.999 mV, 0.000 to 10.250 V dc or 0.00
to 60.00 V dc.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select V for the FUNCTION and 10V, 1V, 60V or 100
mV for the RANGE.

Connect the red input lead (+) of the PIE 820-ELITE to the more
positive point of the break and the black input to the more
negative point.

Signals above the maximum scale are limited by protection
circuitry with “OVERRANGE” flashed on the display and the red
OVERLOAD LED lit.

The PIE 820-ELITE measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch

q

to MAX to see the highest reading and to

MIN to see the lowest reading. Press and hold the knob

e

to

clear the MAX and MIN readings.

Read mV