Specifications – Practical Instrument Electronics 820-ELITE User Manual

Page 34

Advertising
background image

Page 32

Specifications

General
Operating Temp Range

-20 to 60 °C (-5 to 140 °F)

Storage Temp Range

-30 to 60 °C (-22 to 140 °F)

Temperature effect

≤ ± 0.01 %/°C of Full Scale

Relative Humidity

Range

10 % ≤RH ≤90 % (0 to 35 °C),

Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C),

Non-condensing

Normal Mode Rejection 50/60 Hz, 50 dB
Common Mode Rejection 50/60 Hz, 120 dB
Noise

≤ ± ½ Least Significant Digit from 0.1 to 10 Hz

Size

5.63x3.00x1.60” 143x76x41mm (LxWxH)

Weight

12.1 ounces, 0.34 kg with boot & batteries

Batteries

Four “AA” Alkaline 1.5V (LR6)

Battery life

Read Functions: ≥ 20 hours
Source mA: ≥ 14 hours @ 12 mA into 250Ω
Pwr/Meas mA: ≥ 12 hours at 20 mA
Source V, Ω, T/C, RTD & Hz: ≥ 20 hours

Low Battery

Low battery indication with nominal 1

hour of operation left

Protection against

misconnection

Over-voltage protection to 60 vrms

(rated for 30 seconds)
Red LED indicates OVERLOAD or out

of range conditions

Display

High contrast graphic liquid crystal display

with 0.315” (8.0 mm) high digits. LED

backlighting.