RKI Instruments Eagle 2 User Manual

Page 100

Advertising
background image

96 • Detail Settings Button

from Table 5 on page 102 or through testing as described in
“Obtaining a Relative Response Factor” on page 101. A valid
character for the Factor is a value between 0.01 and 25.00.

There are limitations for the full scale and increment values that
depend on the factor for both the high range and the low range.
Table 2 and Table 3 below list these limitations.

Table 2: High Range PID

Factor

Full Scale

(ppm)

Increment

0.25-0.36

1000

1

0.37-0.49

1500

1

0.50-0.61

2000

1

0.62-0.74

2500

1

0.75-0.99

3000

2

1.00-1.24

4000

2

1.25-2.49

5000

2

2.50-3.74

10000

10

3.75-4.99

15000

10

5.00-7.49

20000

10

7.50-9.99

30000

20

10.00-
12.49

40000

20

12.50

50000

2500

Table 3: Low Range PID

Factor

Full Scale

(ppm)

Increment

0.20-0.29

10.00

0.01

0.30-0.39

15.00

0.01

0.40-0.49

20.00

0.01

Advertising