70 • Viewing, Printing, and Deleting Data in the Last Calibration Window

Before — shows the settings prior to calibration.

After — shows the settings after calibration.

A. Cal. — lists the auto-calibration setting for each channel
of the Eagle 2. If a Eagle 2 passes its calibration, the
“After” column should match the “A. Cal.” column. If the
Eagle 2 fails calibration on any of its channels, those
channels will retain the previous calibration information.

NOTE: If a unit is calibrated using Single Calibration in the

Eagle 2’s Calibration Mode (see the Eagle 2 Operator’s
Manual) it is possible for the “After” reading to be
different from the “A.Cal” setting if the unit was set to a
level different than the “A.Cal” setting.

Cal. Due (Days) — shows when calibration is due in days
(e.g., “Remaining 25 Day” means that calibration is due in
25 days, and “Now” means that calibration is due
immediately).

It is not possible to print any information when the Calibration
Record view option is selected.

To view and print all past calibrations for an instrument, see
“Calibration History” on page 26.

Deleting Last Calibration Data

To delete an instrument and its calibration data from the Last
Calibration Window:

1. With the program launched, click Last Calibration on the

right side of the program window. The Last Calibration
Window will appear.

2. Select the Need Calibration or Calibration Date option.

3. Click the first row for the instrument whose calibration

information you want to delete to select it. It will be
highlighted to show that it is selected.