Overwriting Data in the Eagle 2 • 21

Overwriting Data in the Eagle 2

The Eagle 2’s Data Log Overwrite function is factory set to On
so that when the Eagle 2’s data logging memory becomes full, it
begins to overwrite the oldest interval trend data with new
internal trend data. Download data regularly to avoid over-
writing data in the Eagle 2 before it can be downloaded. The
Data Log Overwrite function is accessible using the Eagle 2
Setup Mode. To set the Data Log Overwrite function to Off, see
the Eagle 2 Operator’s Manual. When the Data Log Overwrite
function is set to off, the Eagle 2 will stop saving data when its
data logging memory is full.

The Data Log Overwrite function applies only to interval trend
data. All other data, such as alarm trend data, event data, or
calibration data, will continue to be saved when the memory is
full. If the maximum number of each of these types of data has
been reached, the oldest data will be overwritten.