RKI Instruments Eagle 2 User Manual

Page 99

Advertising
background image

Detail Settings Button • 95

Full Scale
This is the full scale value for the target gas.

Point
The point value indicates to what decimal place the gas
readings are shown. A value of 1/1 indicates a reading to the
“ones” place while a value of 1/100 indicates a reading to the
“hundredths” place.

Unit
The unit describes what units the gas reading is provided in.
All units are in ppm. While this parameter may be changed in
the csv file, units of ppm are the only acceptable units and
changing this parameter to %LEL or %vol will change the unit
box to red.

Digit
The digit is the increment of the gas readings.

1st
This column is for the low alarm point of each gas in ppm
units.

2nd
This column is for the high alarm point of each gas in ppm
units.

STEL
The STEL column displays the STEL values for each gas.

TWA
The TWA column displays the TWA values for each gas.

AutoCal
The AutoCal values are those that come up during the auto
calibration procedure. They are default values that may be
changed if the gas concentration in the calibration cylinder is
different.

Factor
This value is the Relative Response Factor for the PID
channel. The factor for each pre-defined gas is factory
defined. The factor for the user defined gas must be obtained

Advertising