30 • Viewing, Printing, Exporting, and Deleting Data in the Data Window

After specifying the file name, file type, and file location click
the Save button to save the file to the specified location.

12.To go back and view other data, click the Return button in

the upper right corner of the Data Window or the Data
control button on the right side of the program window and
select the data you want to view.

Event Data

The Eagle 2 not only saves trend files which include logged data
at scheduled times, but also saves the 100 most recent alarm
events and the 100 most recent trouble events. When an
instrument is downloaded, the Eagle 2 Data Logger
Management Program will retrieve these events from an
instrument and save them in alarm event files and trouble event
files for each instrument that is downloaded.

Alarm event files save the time, instrument channel, gas, and
alarm type of every gas alarm event that occurs on a particular
instrument. Warning (low alarm), Alarm (high alarm), STEL,
TWA, and overscale events are saved.

Trouble event files note the time, instrument channel, whether
the event is an instrument system failure or sensor failure and
the specific type of failure. Calibration failures, dead battery
alarms, and sensor failures are among the trouble events that
are saved.