Changing Eagle 2 Instrument Parameters • 77

13.Connect the regulator to the Eagle 2 probe tube using the

sample tubing provided with the calibration kit.

14.Allow the gas to flow for one minute. The current gas

readings will be shown in the Current Reading column.

15.Click the Span button. The program will make the span

adjustments.

16.Disconnect the tubing from the Eagle 2’s probe tube.

17.Remove the regulator from the cylinder.

18.Allow the current readings to return to normal and click the

Exit Calibration button to return to the Eagle 2 Status frame.
The Eagle 2’s pump will turn off.

19.If you wish to use other functions in the software, click one of

the control buttons.

20.If you wish to exit the software, click the Download button to

return to the Download window. Turn off the Eagle 2 by using
the POWER ENTER RESET button on the Eagle 2 or by
clicking the Power Off button and confirming that you want
to turn off the Eagle 2 when the Power Off window appears.
Then click the Exit button to exit the program.

Changing Eagle 2 Instrument Parameters

To make changes to parameters stored in the Eagle 2, use the
Eagle 2’s Data Logger Management Program’s Set Window.
Follow the steps below to make these changes.

1. Launch the Eagle 2 Data Logger Management Program.

2. If the Automatic Download selection box is selected,

deselect it.

3. Connect an Eagle 2 to the program. See “Downloading Data

from the Eagle 2” on page 15, instruction 2 through
instruction 3, to connect an instrument to the program.

4. Once the connection is made, the Complete Download,

Instrument Information, Clear Logger Data, and Power