Calibration history – RKI Instruments Eagle 2 User Manual

Page 30

Advertising
background image

26 • Viewing, Printing, Exporting, and Deleting Data in the Data Window

Calibration History

The Eagle 2 is capable of saving calibration information for up to
the 100 most recent calibrations. This calibration history is
retrieved by the Data Logger Management Program when data
is downloaded from the Eagle 2 using either the Complete
Download
download command or the Download Logger Data
download command.

Instrument calibration information is also available in the Last
Calibration Window. The information regarding the most recent
successful calibration for each downloaded instrument along
with whether or not an instrument is due for calibration can be
accessed using the Last Calibration button on the right side of
the program window. See “Viewing, Printing, and Deleting Data
in the Last Calibration Window” on page 66 for a complete
description of the information that can be accessed by the Last
Calibration
button.

All downloaded calibration information for all downloaded
instruments is available in the Calibration History folders in the
Data Window. This information is saved in a Calibration History
Folder that is located in an untitled folder for each instrument.
The calibration history files are differentiated by instrument. The
calibration information available here is more comprehensive
than that in the Last Calibration Window. The calibration
information for all calibrations downloaded, whether successful
or not, is saved instead of just the most recent successful
calibration for each instrument.

Advertising