Viewing, Printing, Exporting, and Deleting Data in the Data Window • 43

If there are more than two consecutive data points with the
same readings for all channels, only the first and last of
these consecutive data points are shown.

All events, such as gas alarms or sensor failures, are
shown.

For any event, the data point before and after the event is
always shown.

11.To view the data in graph format, click the Graph selection

button. Five or more scheduled data points are required in
an interval trend data file to be able to display it in graph
format.

NOTE: The data count shown when you have selected an

interval trend data file, as in Figure 26 above, can be
more than five if you have events, such as the Eagle 2
going into and out of alarm, but you may not have five
scheduled data points.