7.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

92

■■■■■■■■

848 Titrino plus

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
off
No evaluation takes place.

for Window = on
Selection

first | greatest | last

Default value

first

first
Only the first equivalence point will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.

Fixed EP1 at

The associated volume will be interpolated from the measuring point list
for the measured value entered. The fixed endpoint must lie between the
first and the final entry in the measuring point list.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–1250.0 - 1250.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–125.00 - 125.00 µA

Selection

off

Default value

off

Fixed EP2 at

See Fixed EP1 at.