5 Operation

848 Titrino plus

■■■■■■■■

39

In the function bar, select Edit and press [OK].

4 Editing the sample data

Edit the sample data and apply with Accept or [BACK].

Note

In addition to the sample data, the method can also be modified,
except in cases where the determination is running.

5 Displaying the live dialog

Select the menu item Live dialog in the main menu and press
[OK].

or

Press [BACK] in the main dialog.

The live dialog is displayed once again.

Editing the sample data of the running determination
When you use the sample table, the editing of the sample data of the run-
ning determination proceeds as described in chapter 5.9.1, page 37. In
addition, you have the option of editing these in the sample table. The
first line always contains the sample data of the running determination.
Simply select for this purpose the Sample table(see "Editing the sample
table", page 38)
menu item in the main menu.

5.9.3

Editing the live parameters

Certain method parameters can be edited while a determination is being
carried out. The only parameters that can be modified are those that can
be selected. Nevertheless, all of the parameters are visible. The modified
parameters are taken into account at once. If you modify, for instance,
the start conditions after the start volume has been dosed, then these
modifications will not be taken into account until the next determination.

Proceed as follows to edit the parameters:

1 Displaying the main dialog

Press [BACK].

The main dialog is displayed. The determination continues to run in
the background.