3 stirring, Stirring – Metrohm 848 Titrino plus User Manual

Page 60

Advertising
background image

5.13 Manual Control

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

848 Titrino plus

The status changes to busy. The current measured value and the
measuring temperature are displayed.

5 Stop the measurement

Press [STOP] or [BACK].

The measurement is stopped. The status changes back again to
ready. If you stop the measurement with [BACK], then the dialog
will also be exited.

5.13.3

Stirring

You can control a connected stirrer manually.

Proceed as follows:

1 Open the manual control

In the main dialog, select Menu and press [OK].
The main menu opens.

Select the menu item Manual control and press [OK].

Manual control opens.

2 Set the stirring rate

Select the Stirrer item.

In the function bar, select Stir– or Stir+.
The stirring rate will be increased or decreased by one step each
time the [OK] key is pressed.
The algebraic sign changes the direction in which the stirring is
done. When the stirrer is viewed from above, this means:

– "+": counterclockwise rotation
– "–": clockwise rotation