Metrohm 848 Titrino plus User Manual
Page 58
5.13 Manual Control
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
848 Titrino plus
2 Select the dosing function
■
Select the Dosing item.
■
In the function bar, select ADD and press [OK].
3 Configure the dosing function
Note
■
The dosing and filling rate should be decreased for viscous liq-
uids.
■
The maximum dosing and filling rate depends on the cylinder
volume (see Chapter 10.1.1, page 117) .
■
Enter the desired volume.
■
Enter the dosing rate.
■
Enter the filling rate.
4 Start dosing
■
Press [START].
The status changes to busy, the dosed volume is displayed. When a
cylinder volume has been added, the dosing cylinder will be refilled
automatically.
5 Fill the cylinder
■
Press [STOP] or [BACK].
The dosing cylinder is filled. If you start the filling with [BACK], then
the dialog will also be exited.