9.2 Miscellaneous

■■■■■■■■■■■■■■■■■■■■■■

116

■■■■■■■■

848 Titrino plus

Problem

Cause

Remedy

The electrode responds too
slowly.

Replace the electrode.

9.2

Miscellaneous

Problem

Cause

Remedy

No report is printed.

The printer is not recog-
nized by the instrument.

Switch the 848 Titrino plus off and on
again.

Use a hub on the USB connector and con-
nect the printer to the hub.

The printer model is not
compatible.

Use a printer which fulfills the required specifi-
cations (see Chapter 10.4.4, page 121).

The USB keyboard
or the PC mouse
does not function.

The keyboard or mouse is
not recognized by the
instrument.

Switch the 848 Titrino plus off and on
again.

Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.

The keyboard or the mouse
is not compatible.

Use a model which fulfills the required specifi-
cations (see Chapter 10.4, page 119).

The display is no
longer readable.

The contrast is set incor-
rectly.

Adjust the contrast correctly (see Chapter 6.1,
page 52)
.

Mettler XP balances
send "R" or "O ----"
as ID1.

The automatic calibration
of the balance is switched
on.

Switch off the automatic calibration.

Message 020-511
"Action not possi-
ble" is displayed.

The USB stick is no longer
connected.

1. Connect the USB stick.
2. Switch the instrument off and on again.

The USB stick is full.

Use a different USB stick.

Delete files with the aid of a PC.