3.7 Robotic arms with beaker sensor

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

855 Robotic Titrosampler

3 Screw the robotic arm to the Swing Head tightly with the screws and

washers provided.

3.7

Robotic arms with beaker sensor

For safety reasons, the presence of a beaker on the sample rack of a Sample
Processor can be detected. Some robotic arm model versions are therefore
equipped with a beaker sensor.

6.1462.150

1

2

3

Figure 14

Connecting a beaker sensor (for example 6.1462.150)

1

Plug of the connection cable

2

Connection socket on the 786 Swing
Head

3

Beaker sensor
Contact sensor in accordance with the Piezo
principle

A robotic arm with beaker sensor is mounted as described on page 18. The
connection of the sensor cable must take place while the instrument is
switched off.

The beaker sensor is automatically recognized when switching on the instru-
ment.