7.1 Measuring interface

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

855 Robotic Titrosampler

7 Technical specifications

7.1

Measuring interface

The 855 Robotic Titrosampler has one galvanically isolated measuring inter-
face.

The measuring cycle is 100 ms for all measuring modes.

7.1.1

Potentiometry

One high-ohm measuring input (Ind.) for pH, redox or ISE electrodes and
one reference measuring input (Ref.) for a separate reference electrode.

Input resistance

>1· 10

12

Ω

Offset current

<1· 10

-12

A (under reference conditions)

Measuring mode

Measuring
range

-20…+20 pH

Resolution

0.001 pH
0.1 mV

Measuring accu-
racy

±0.003 pH
±0.2 mV
(±1 digit, without sensor error, under reference conditions)

7.1.2

Polarizer

A measuring input (Pol.) for a polarizable electrode.

Measuring mode
Ipol

Determination with adjustable polarization current.

Polarization cur-
rent

–125.0…+125.0 µA (in increments of 2.5 µA)
–125…–121 µA / +121…+125 µA: non-guaranteed values, dependent
on reference voltage +2.5 V

Measuring
range

-1200…+1200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

Measuring mode
Upol

Determination with adjustable polarization voltage

Polarization
voltage

-1250…+1250 mV (in increments of 25 mV)