1 connecting dosing devices, Connecting dosing devices, Figure 23 – Metrohm 855 Robotic Titrosampler User Manual

Page 38: Connecting a dosing device

Advertising
background image

3.13 Connecting MSB devices

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

855 Robotic Titrosampler

3.13.1

Connecting dosing devices

Three dosing devices can be connected to the instrument.

The types of dosing devices that are supported are:

800 Dosino

700 Dosino

805 Dosimat

685 Dosimat

Warning

If a Dosino is connected to the 855 Robotic Titrosampler then the con-
nection cable must be equipped with a T.2400.102 ferrite core. The ferrite
core reduces any interference voltages that may occur and thus ensures
compliance with strict EMC standards pursuant to applicable technical
norms, see Chapter "Technical Data".

Proceed as follows:

1 Mounting ferrite core

Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.

2 Connect a dosing device

Exit the control software.

Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.

Start the control software.

USB 1

Contr.

MSB 2

MSB 3

T.2400.102

Figure 23

Connecting a dosing device