3 Installation

855 Robotic Titrosampler

■■■■■■■■

35

Proceed as follows:

1 Insert the enclosed countersunk screw from below into the opening on

the very back of the base plate.

2 Place the washer (flat side upwards) over the countersunk screw and

tighten the support rod with the countersunk screw. The necessary
hexagon key is enclosed with the 855 Robotic Titrosampler .

3 Hang the base plate to the assembly rail of the turntable with the hold-

ing clamps. Slide the whole stand plate as near to the tower as possible.
Fix the holding clamps with a hexagon key to the assembly rail.

Note

Before the stand plate is fixed to the tower, e.g. a magnetic stirrer
or a measuring cell can be mounted on the support rod.

4 Guide the bracing from above over the support rod.

5 Loosen the screw for fastening the bracing on the rear panel of the

tower (see figure) and fix the bracing with the enclosed hexagon screw.

6 Fix the bracing to the support rod with a hexagon key, see figure.

3.16

Mounting the drip pan

Serious damage to the instrument or a danger to the user can occur if
chemicals or liquid samples are spilled. The use of the drip pan (6.2711.060)
is recommended in order to avoid such incidents.

Mounting the drip pan

1

2

Figure 29

Installing the drip pan

Advertising