Metrohm 886 Prof.Thermostat/Reactor User Manual

Page 24

Advertising
background image

3.3 Mounting base tray and bottle holder (optional)

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

1

2

3

1 Attach the bottle holder onto the topmost instrument so that the

openings in the bottle holder exactly match the screw threads on the
top surface of the instrument.

2 Insert the two cylinder screws and tighten them with the 3 mm hexa-

gon key.

3 Insert covering stoppers .

After having attached the bottle holder, all connections which were
undone at the beginning of the process, must be reconnected. Proceed as
follows:

Restoring the loosened connections

1 Connect the USB cable.

2 Connect the MSB cable.

3 Plug in the mains cable.

4 Reconnect drainage tubings (see manual of the IC instrument).

Possibly, a longer section of silicone tubing 6.186.020 must be cut to
fit and mounted (see also the manual for the IC instrument).

5 If one of the instruments in the stack is equipped with a leak sensor

connection socket, connect the leak sensor (see manual of the IC
instrument)
.

Advertising