3 Installation

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

■■■■■■■■

29

4 Mount the Y connector

6.2744.330 Y connectors are used to add the reagent. These are
supplied with each 6.2845.200 reactor. You can attach several Y
connectors to the reactor block.

Attach the Y connectors to the reactors using the included
screws.

3.8.2

2.886.0210 Professional Thermostat

Equipped with the 6.2845.600 column holder, the 886 Professional Ther-
mostat / Reactor provides the functionality of a column thermostat.

The easiest way to connect capillaries is outside the oven.

All capillaries must be fed into or out of the oven through the capillary
feed-throughs at the right and left side of the oven door.

If you have placed the 886 Professional Thermostat / Reactor under or
next to the IC instrument, proceed as follows:

1.

Feeding capillaries into the oven

2.

Connect the capillaries with the columns

3.

Insert column holder

Start with "Feeding capillaries into the oven", page 24.

If you have placed the 886 Professional Thermostat / Reactor on top of
the IC instrument, proceed as follows:

1.

Connect the capillaries with the columns

2.

Insert column plate

3.

Guide capillaries out of the oven

Start with "Connect the capillaries", page .24

Note

In order to keep the dead volume as low as possible, make sure that
capillary connections are as short as possible.

Always shorten capillaries with the 6.2621.080 capillary cutter to obtain
flawless level cut capillary ends.