3.8 Installing accessories

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

1

2

3

4

1 Lower the latches

Loosen the outer screws of both latches with the 6.2621.140 hexa-
gon key 2.5 mm turning them counterclockwise.

The latches open and unlock the insertion rails of the heating plate.

2 Insert the column plate

Insert the column plate into the oven by fitting the two bolts at
the sides of the column plate into the insertion rails of the heating
plate.

Slide the column plate into the oven as far as it will go. Take care
not to jam or pinch the capillaries.

3 Lift the latches

Tighten the outer screws of both latches with the 6.2621.140 hexa-
gon key 2.5 mm turning them clockwise.

The latches lift and fix the column holder on the heating plate.

4 Connect the capillaries to the system

In case the free capillary ends are not yet guided through the capil-
lary feed-throughs, guide them out of the oven.

Connect the free end of the preheating capillary with the injection
valve.