3.8 Installing accessories
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)
1
2
3
4
1 Lower the latches
Loosen the outer screws of both latches with the 6.2621.140 hexa-
gon key 2.5 mm turning them counterclockwise.
The latches open and unlock the insertion rails of the heating plate.
2 Insert the column plate
■
Insert the column plate into the oven by fitting the two bolts at
the sides of the column plate into the insertion rails of the heating
plate.
■
Slide the column plate into the oven as far as it will go. Take care
not to jam or pinch the capillaries.
3 Lift the latches
Tighten the outer screws of both latches with the 6.2621.140 hexa-
gon key 2.5 mm turning them clockwise.
The latches lift and fix the column holder on the heating plate.
4 Connect the capillaries to the system
In case the free capillary ends are not yet guided through the capil-
lary feed-throughs, guide them out of the oven.
■
Connect the free end of the preheating capillary with the injection
valve.