3.8 Installing accessories

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

Optional: attaching additional reactors

To attach additional reactors, you need the following accessories:

6.2845.200 Reactor complete

6.2621.140 hexagon key 2.5 mm

1

6.2845.200

6.2845.200

2

6.2845.200

6.2845.200

6.2845.200

6.2845.200

3

6.2845.200

6.2845.200

6.2845.200

6.2845.200

4

6.2845.200

6.2845.200

6.2845.200

6.2845.200

1 Attach two reactors

The originally mounted reactor is attached to the reactor plate with
four screws turned in from the bottom of the plate.

Remove the four screws using the 2.5 mm hexagon key.

Reattach the reactor at the far end of the reactor plate.

Attach a second reactor from the 6.2845.200 accessory set at the
front end of the reactor plate using the the included screws and
the 2.5 mm hexagon key.

2 Attach additional reactors

Place two additional reactors on top of the other two.

3 Fasten the reactors

Fasten each reactor on both sides to the reactor below using the
included clips.