Figure 15, Connecting the columns – Metrohm 886 Prof.Thermostat/Reactor User Manual

Page 39

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

■■■■■■■■

31

1

2

3

4

5

Figure 15

Connecting the columns

1

Main column
iColumn clicked into the column holder.
Maximum length: 150 mm.

2

Auxiliary column
Either a second column (of 100 mm length)
with on column guard system or guard col-
umn only.

3

Preheating capillary
Part of accessories set 6.2845.600.

4

Capillary
Connection between main column and aux-
iliary column.

5

Capillary
Connection between main column and
detector.

Connecting and inserting columns

To install the columns, you need:

main column: an intelligent separation column (iColumn) of maximum
150 mm length (e.g. Metrosep A Supp 15 -150/4.0)

auxiliary column: a guard column compatible with the main column
(e.g. Metrosep A Supp 15 S-Guard/4.0) or
a second column compatible with the main column (e.g. Metrosep A
Supp 15 - 100/4.0) with an on column guard column (Metrosep A
Supp 15 Guard/4.0)

sections of the included capillary 6.1831.100

the capillary cutter (6.2621.080)