5 housing, 6 reactor / column thermostat, Housing – Metrohm 886 Prof.Thermostat/Reactor User Manual

Page 51: Reactor / column thermostat

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

9 Technical specifications

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

■■■■■■■■

43

9.5

Housing

Dimensions

Width

365 mm

Height

131 mm

Depth

380 mm

Material of base
tray, housing and
covering plate

Polyurethane hard foam (PUR) with flame retardation

Operating ele-
ments

Indicators

LED for power display

On/Off switch

On the rear of the instrument

9.6

Reactor / Column thermostat

Number of reac-
tors

up to 4

Number of col-
umns

up to 2

Adjustable tem-
perature range

0 °C…+150 °C, adjustable in steps of 0.1 °C

Stability

less than 0.05 °C deviation

Reproducibility

better than ± 0.2 °C

Temperature
accuracy

< 1 °C (typical)

Heating up time

< 10 Min. from 20 °C to 40 °C

Maximum pres-
sure

Reactor

2 MPa

Heating type

Resistor type thermostat for intelligent reactors and intelligent separa-
tion columns

safety shutdown

Function

Automatic switch off when maximum temperature is exceeded

Maximum tem-
perature

nominal 169 °C, response time: immediately

Advertising