3 keyboard connection, Keyboard connection, Fig. 5 – Metrohm 824 Easy Sample Changer User Manual

Page 16: Fuse holder, Fig. 6, Rear panel of instrument

Advertising
background image

2.3 Keyboard connection

10

824 Easy Sample Changer,

Installation

Pull out mains cable

Remove the mains cable from the mains supply connection of the 824

Easy Sample Changer.

Remove the fuse holder

Use a screwdriver to loosen the fuse holder beside the mains supply

connection and remove it completely.

Check the fuse and replace it

Carefully remove the built-in fuse for the intended voltage from the fuse

holder and check its specifications (the position of the fuse in the fuse

holder is indicated by the white arrow beside the voltage range):

100

120 V 0.5 A (slow blow)

Metrohm-No. U.600.0014

220

240 V 0.25 A (slow blow)

Metrohm No. U.600.0011

Insert fuse

Exchange the fuse if necessary and replace it in the fuse holder.

Insert fuse holder

Depending on the required mains voltage, insert the fuse holder so that

the white arrow of the corresponding voltage range points to the white

bar; this is to the right of the fuse holder (see below).

220

- 24

0

V

100
- 12
0

V

220
- 240
V

100

- 12

0

V

Fig. 5 Fuse holder

Attach connections cover

Position the connections cover in the guide rail above the connections

strip. The connections cover protects the connections from spilt chemi-

cals.

2.3 Keyboard connection

Type 1.824.0010 Nr.

0010/01104

Made by Metrohm Herisau Switzerland

RS 232

Keyboard

Fig. 6 Rear panel of instrument

The keyboard is attached to the connection 24

provided for it on the rear panel of the instrument. To

remove the connection press both sides of it together.

100 – 120 V

220 – 240 V

Mains switch

Mains connection

Fuse holder

Advertising