3.2 Settings

22

824 Easy Sample Changer,

Operation

3.2 Settings

In order to be able to work comfortably with the 824 Easy Sample

Changer a few settings must first be made.

The various lift settings depend on the rack used; this means that

they must be made for each sample rack according to the size of the

sample vessels used.

The stirrer rate is method-specific and is stored in the active method.

If you use different methods then the stirrer speed must be optimized

separately for each method.



3x

HOLD

SELECT

PUMP/STIR

Method

Config

SELECT

Place a sample rack on the turntable and press <STOP/RESET>.

Place a filled sample beaker at the rack position in front of the lift in

order to make the following settings.

Press the <SELECT> key three times.

The SELECT LEDs show that the 824 Easy Sample Changer is now in

the configuration mode.

You can now use the <Í> and <Î> keys to switch between the four

possible settings in this mode.

3.2.1 Setting the working position of the lift

STIR 1

STIR 2

The lift with the titration head can be easily moved up and down with

the <Ï> and <Ð> keys respectively.

Set the working height of the lift so that the electrode is immersed and

the stirrer can work efficiently.

Accept/ continue with <Î>.

3.2.2 Setting the rinsing position of the lift

STIR 1

STIR 2

The lift with the titration head can be easily moved up and down with the

<Ï> and <Ð> keys respectively.


If standard method 4 is used then the rinsing position of the lift is

required for aspirating the used sample. Set the rinsing position of the

lift so that the sample vessel can be emptied completely. A rinsing

position which is lower than the working height can be selected. The

mounting position of the aspiration tubing can be selected as required.
Accept/ continue with <Î>.