3.4 Automatic operation

824 Easy Sample Changer,

Operation

25

3.4 Automatic operation

3.4.1 Preparing a sample series

Prepare all the samples to be processed. Select the sample vessel

size according to your type of sample rack. The setting for the pre-

defined lift position must be matched to the size of your sample

vessels.

Place the sample vessels on the rack. Start with rack position 1.

Do not forget to place a rinsing/conditioning beaker at the highest

rack position. Most sample changer methods require a special

beaker, see method description p. 29ff.

In order to process a series of samples a suitable method must first be

loaded, see below.

3.4.2 Method selection

1x <SELECT>

STIR 1

STIR 2

3x <SELECT>

After pressing the <SELECT> key a method can be selected by using

the <Ð> and <Ï> keys. Four predefined simple standard methods are

available (Methods 1,2,3 and 4).

LEDs LED 1 to 4 show the selected method. The selected method is

accepted with <SELECT>.

Return to the normal operating mode with 2x <SELECT>.

3.4.3 Starting a method

START

A method is started with the <START> key.

When a method is started the sample changer is first initialized, i.e.

the lift is moved upward as far as it will go and the sample rack is

moved to the starting position. The magnet code of the rack is read

in and the corresponding rack table, in which the beaker positions

on the rack are defined, is loaded.

While the method is running the status LED blinks slowly (once per

second).

After the start of the method the starting sequence is carried out

once. This contains commands for preparing for a series of sam-

ples.

When all rack positions have been processed the final sequence is

carried out.