8.5 Sample beakers for Karl Fischer titrations

824 Easy Sample Changer,

Annex

49

8.4.3 Aspiration tip

The Metrohm 772 Pump Unit with aspiration equipment can be used for

aspirating used sample solutions without any problems. If the sample

solutions contain solids, precipitates (e.g. silver chloride) or sticky sub-

stances then a peristaltic pump such as the Metrohm 772 Pump Unit

must be used, as diaphragm pumps can easily be blocked up or stuck

together by solids.

The 6.1543.170 Aspiration tips can be used with all titration heads with

SGJ9 openings, see above drawing.

Take the chemical resistance of the tubing and connections into ac-

count.

Polyethylene canisters (Metrohm Order. no. 6.1621.000) can be used

as waste canisters for aqueous media.

The 6.1821.000 Aspiration Tube is used with the KFT titration head. It

should always be used as it ensures trouble-free penetration of the

aluminum foil on the KF sample vessels.

8.5 Sample beakers for Karl Fischer titrations

For Karl Fischer titrations the

measured-out sample is placed in a

6.1432.210 Sample Beaker together

with a 6.1903.010 Stirrer Bar. The

beaker is then covered centrally

with 6.2820.000 Aluminum foil which

is kept in place by the foil holder.

Fig. 26

Karl Fischer sample beaker

During the method sequence the lift with the titration head is moved to

the working position. The aspiration tube should be the first object to

penetrate the Al foil.

Attention

Always attach the 6.1821.000 Aspiration Tube to the KFT titration

head, even when you do not intend to aspirate sample solutions. The

aspiration tube is used to penetrate the Al foil.

Use only 6.2820.000 Al Foil as this has a defined thickness

(0.010 mm).

Sample

beaker

Al foil

Foil holder

6.2037.040