2.7 Serial connection (RS232)

18

824 Easy Sample Changer,

Installation

2.7 Serial connection (RS232)

A personal computer can be connected to the serial RS232 interface

25. In this way the 824 Easy Sample Changer can be remotely con-

trolled (in the same way as the Metrohm 730 Sample Changer), pro-

vided that a suitable software such as Tinet 2.5 is installed.

824

Fig. 18

Connecting a computer

A requirement for correct data transfer is that the transmission parame-

ters are correctly set; these must correspond to the settings at the inter-

face of the connected device.

The standard transmission parameters for the 824 Easy Sample

Changer are:

Baud Rate:

9600 Baud

Data Bit.

8

Stop Bit:

1

Parity: none

Handshake: HWs

Cable 6.2125.060 (25p/25p)

oder 6.2125.110 (25p/9p)