7 troubleshooting, 1 error messages, Troubleshooting – Metrohm 824 Easy Sample Changer User Manual

Page 43: Error messages

Advertising
background image

7.1 Error messages

824 Easy Sample Changer,

Troubleshooting

37

7 Troubleshooting

7.1 Error messages

If a fault or error occurs then LEDs 1 to 4 start to blink in particular

combinations (binary error code). The status LED goes out.

HOLD

SELECT

<START>

or

<STOP>

Instrument error messages (blinking LEDs 1…4) must be acknowl-

edged with the <HOLD> key. The 824 Easy Sample Changer then

enters the HOLD mode.

When the fault has been remedied then the run can be continued with

the <START> key or canceled with the <STOP> key.

LED 1…4

Description / Remedy

Error 1:

Beaker missing

After a rack rotation no beaker is found at the moved-to position.
Remedy: Press <HOLD>. Position beaker and press <START> or

cancel sample series with <STOP>.

Error 2:

Lift not in rotating position

The sample rack must be at or above the predefined rotating position

before it can rotate.
Remedy: Press <HOLD> and <STOP>. Move lift upward with <Ï>

and then rotate rack again with <Í> or <Î>.

Error 3:

Sample rack error

An incorrect or no sample rack at all is in position. Possibly the magnet

code of the rack cannot be read or the instrument does not recognize

the magnet code.
Press <HOLD>. Position Metrohm standard sample rack. Press

<STOP> or <START>.

Error 4:

Invalid rack position

An attempt has been made to move to an invalid rack position (method

error).
Remedy: Press <HOLD> and <STOP>. Check method.

Error 5:

Sample changer not ready

The 824 Easy Sample Changer cannot carry out the command as it is

busy doing something else.
Remedy: Press <HOLD> and <STOP>. Wait and then repeat com-

mand.

1
2
3
4

1
2
3
4

1
2
3
4

1
2
3
4

1
2
3

4