7.1 Error messages

38

824 Easy Sample Changer,

Troubleshooting

Error 6:



Sample rack and swing head not compatible

The positioned sample rack cannot be used together with the swing

head and vice versa. Multi-row sample racks must be used with the

swing head, single row racks without the swing head.
Remedy: Press <HOLD> and <STOP>. Change rack and press

<STOP/RESET>.

Error 10:

Mains overload

The power supply cannot cope with the simultaneous use of all

components (stirrers, pumps, lift).
Remedy: Press <HOLD> and <STOP>. Switch off a stirrer or a pump.

If necessary, switch the instrument off and on again.

Error 11:


Method store full

The memory for the user-defined methods is full.
Remedy: Switch the instrument off and on again. Delete any methods

which are no longer required from the memory (via RS232 interface or

with the "6.2142.010 SC Controller" keypad).

Error 12:

Sample changer overload

Load or resistance is too large to carry out the selected action.
Remedy: Press <HOLD> and <STOP>. Remove the mechanical

hindrance to the lift or sample rack.

Error 13:

RS232 error, interface error

Error during data transfer or control by a PC.
Remedy: Press <HOLD> and <STOP>. Observe the PC software

display. Check the interface parameters. If necessary, consult the PC

software manual.

Error 14:

Exceptional instrument error

An exceptional fault has occurred in the 824 Easy Sample Changer

hardware.
Remedy: Switch the instrument off and on again. If the fault occurs

again please contact the service department of your local Metrohm

agency.

Error 15:

Program stopped (trap error)

An unexpected and non-recoverable program error has occurred.
Remedy: Switch the instrument off and on again. If the fault occurs

again please contact the service department of your local Metrohm

agency.