8.2 Method listings

42

824 Easy Sample Changer,

Annex

8.2 Method listings

8.2.1 Method 1

824 Easy Sample Chan.

5.824.0010

¬

Report header with program version

Parameters

method

1

¬

Method name

number of samples:

rack

¬

No. of samples (whole sample rack)

>start sequence

1 CTL:Rm:

INIT

¬

Initialize remote interface

2 SAMPLE:

-

1

¬

Initialize sample

3 SAMPLE:

+

1

position

>sample sequence

1 MOVE

1

:

sample

¬

Move next sample in front of tower

2 LIFT: 1

:

work mm

¬

Move lift to working position

3 STIR: *

:

ON s

¬

Switch on stirrer

4 CTL:Rm:

START dos*

¬

Start addition

5 WAIT

10 s

¬

Waiting time

6 CTL:Rm:

START device1

¬

Start instrument/titrator

7 SCN:Rm

:

End1

¬

Wait for end of determination [EOD]

8 STIR: *

:

OFF s

¬

Switch off stirrer

9 LIFT: 1

:

shift mm

¬

Move lift to rotating position

10 WAIT

3 s

¬

Waiting time: 3 s for drips

>final sequence

1 MOVE

1

:

spec.1

¬

Move conditioning beaker in front of tower

2 LIFT: 1

:

work mm

¬

Immerse electrode

>changer settings

------ Sample changer settings --------

rack number

0

lift rate 1

25 mm/s

shift rate

20

shift direction:

auto.

beaker test mode:

single

¬

If sample beaker is missing then the

on beaker error:

MOVE

next one will be moved to automatically

>stirring rates

------ Stirring rates ---------------------------

stirrer 1

3

stirrer 2

3

>manual stop

------ Reaction to manual stop --------------

CTL Rmt:

STOP device1

¬

Stop instrument/titrator

CTL RS232: