2 method labels, 3 method 1, Method labels – Metrohm 824 Easy Sample Changer User Manual

Page 36: Method 1

Advertising
background image

4.2 Method labels

30

824 Easy Sample Changer,

Standard methods

4.2 Method labels

On the keyboard covering film you will find a gray marking field with 4

lines. You can enter a short description for each of the four sample

changer methods.

Use a soft pencil or waterproof marker for the labels.

When choosing a marker make sure that the writing can easily be

wiped off with alcohol. Pencil markings can be removed with a soft

rubber.

A brief description of the standard methods of the 824 Easy Sample

Changer is given below.

4.3 Method 1

Use:

For simple titrations (including KF titrations) or pH measurements (with-

out calibration). The initial addition of auxiliary reagents or solvents is

possible.

Required instruments:

Metrohm titrator (Titrino or Titrando) or pH meter

Metrohm Dosimat (optional)

Metrohm 802 Rod Stirrer or 741 Magnetic Stirrer

Features:

No electrode rinsing after sample processing. The electrode is im-

mersed in the conditioning beaker at the end of a series of samples.

Preparation:

Fill a sufficiently large conditioning vessel with a suitable solvent and

place it in the highest rack position.

Procedure:

(Start sequence) Initialize the sample changer

Move to sample and lower lift to working position

Switch on stirrer

If a Dosimat is connected (optional): add auxiliary reagent or sol-

vent and wait until addition is completed

Start titration/measurement and wait until it is completed

Switch off stirrer

Move lift to rotating position and allow electrode to drip

(Final sequence) Move to conditioning beaker and immerse elec-

trode

Advertising