4.1 Information about the methods

824 Easy Sample Changer,

Standard methods

29

4 Standard

methods

On delivery the 824 Easy Sample Changer already contains four opti-

mized standard methods for processing series of samples. Simple titra-

tion tasks can be carried out with these methods without having to

make complicated settings.

4.1 Information about the methods

The Metrohm sample changer methods consist of three different se-

quences for different purposes and a series of specific settings.

Sequences are made up of a series of individual commands with which

the functions of the sample changer can be programmed. For example,

the command 'LIFT 1 work' causes the Lift 1 of a sample changer to

move to the predefined working height.

Metrohm sample changer methods differentiate between three dif-

ferent sequences:

Start sequence

The start sequence is carried out once at the start of a series of

samples.

Sample sequence

The sample sequence is carried out once for each sample.

Final sequence

The final sequence is carried out once at the end of a series of

samples.

sample sequence

start sequence

final sequence

<START>

number
of samples

normal state

Series operating

The number of samples is limited only by the capacity of the sample

rack. In principle, in the standard methods of the 824 Easy Sample

Changer all the positions of a sample rack are processed. Empty rack

positions are automatically jumped over.

The standard methods of the 824 Easy Sample Changer can be used

with all Metrohm standard sample racks. In all these methods the

highest (last) rack position is used for a rinsing or conditioning beaker.

Make sure that a sufficiently large vessel is always placed in this posi-

tion.