Prepare instrument, Calibrate system – Bio-Rad Bio-Plex Pro™ Human Th17 Cytokine Assays User Manual

Page 11

Advertising
background image

9

2. Prepare Instrument

Start up and calibrate the Bio-Plex

®

system with Bio-Plex Manager

software prior to setting up the assay. The calibration kit should be run
daily or before each use of the instrument to standardize the fluorescent
signal. For instructions on using other xMAP system software packages,
contact Bio-Rad Technical Support.

The validation kit should be run monthly to ensure performance of fluidics
and optics systems. Refer to either the software manual or online Help for
directions on how to conduct validation.

Start Up System (Bio-Plex 100, 200, or similar)

1. Empty the waste bottle and fill the sheath fluid bottle before starting

if high throughput fluidics (HTF) are not present. This will prevent
fluidic system backup and potential data loss.


2. Turn on the reader, XY platform, and HTF (if included). Allow the
system to warm up for 30 min (if not already done).

3. Select Start up and follow the instructions. If the system is idle
for 4 hr without acquiring data, the lasers will automatically turn off.
To reset the 4-hr countdown, select Warm up and wait for the
lasers/optics to reach operational temperature.

Calibrate System

1. Select Calibrate and confirm that the default values for CAL1
and CAL2 are the same as the values printed on the bottle of
Bio-Plex calibration beads. Use the Bio-Plex system low RP1
target

value.

2. Select OK and follow the software prompts for step-by-step
instructions for CAL1 and CAL2 calibration.

NOTE!

In Bio-Plex Manager v6.1 and higher, startup, warm up and calibrate

can be performed together by selecting the start up and calibrate icon.