Principle

Technology
The Bio-Plex

®

suspension array system is built around the three core

elements of xMAP technology:

• Fluorescently dyed microspheres (also called beads), each with a distinct

color code or spectral address to permit discrimination of individual tests
within a multiplex suspension. This allows simultaneous detection of more
than 100 different types of molecules in a single well of a 96-well microplate

• A dedicated flow cytometer with two lasers and associated optics to
measure the different molecules bound to the surface of the beads

• A high-speed digital signal processor that efficiently manages the

fluorescence data

Assay Format
Bio-Plex Pro

assays are essentially immunoassays formatted on

magnetic beads. The assay principle is similar to that of a sandwich
ELISA (Figure 1). Capture antibodies directed against the desired
biomarker are covalently coupled to the beads. Coupled beads react
with the sample containing the biomarker of interest. After a series of
washes to remove unbound protein, a biotinylated detection antibody
is added to create a sandwich complex. The final detection complex is
formed with the addition of streptavidin-phycoerythrin (SA-PE) conjugate.
Phycoerythrin serves as a fluorescent indicator, or reporter.

2