Figure 1. Bio-Plex sandwich immunoassay

Data Acquisition and Analysis
Data from the reactions are acquired using a Bio-Plex

®

system or similar

Luminex-based reader. When a multiplex assay suspension is drawn into
the Bio-Plex 200 reader, for example, a red (635 nm) laser illuminates the
fluorescent dyes within each bead to provide bead classification and
thus assay identification. At the same time, a green (532 nm) laser excites
PE to generate a reporter signal, which is detected by a photomultiplier
tube (PMT). A high-speed digital processor manages data output, and
Bio-Plex Manager

software presents data as Median Fluorescence

Intensity (MFI) as well as concentration (pg/mL). The concentration of
analyte bound to each bead is proportional to the MFI of reporter signal.

MAGNETIC BEAD

BIOMARKER

OF INTEREST

CAPTURE

ANTIBODY

STREPTAVIDIN

BIOTINYLATED

DETECTION

ANTIBODY

PHYCOERYTHRIN

REPORTER

FLUORESCENT