Possible Causes

High Background Signal

Incorrect buffer was used
(for example, assay buffer used
to dilute standards)

Accidentally spiked blank wells


Detection antibodies or
streptavidin-PE incubated too long

Poor Recovery

Expired Bio-Plex reagents
were used


Incorrect amounts of components
were added

Microplate shaker set to an
incorrect speed

High end saturation of the
standard curve

Quality controls do not fall within
expected ranges.

Possible Solutions

Use sample matrix standard diluent
to dilute standards.

Be careful when spiking standards.
Do not add any antigens to the
blank wells.

Follow the procedure incubation
time precisely.

Check that reagents have not expired.
Use new or nonexpired components.

Check your calculations and be
careful to add the correct volumes.

Check the microplate shaker speed
and use the recommended setting.
Setting the speed too high may
cause splashing and contamination.
Use the recommended plate shaker.

Make sure that correct shaker speed
and incubation times are used.
Remove S1 for data analysis if needed.

Make sure that the controls are
reconstituted at the same time as
standards, and in the same diuent
(standard diluent HB). Incubate for
precisely 30 min.