4. Prepare Standards and Controls

General Instructions

• It is essential to prepare standards and quality controls (if included)

exactly as described in this section. Incorrect preparation may lead to
low signal or variable measurements from plate to plate

• The product data sheet or peel-off sticker provided with the standards

lists the most concentrated point on the standard curve (S1). Enter this
information into Bio-Plex Manager

software as instructed in Section 8

Using the Quality Controls (optional)

Two-level quality controls (high and low) are included with premixed
Th17 cytokine panels and custom x-Plex kits. Their use is intended for
monitoring the day-to-day quality of assay results.

Selecting a Diluent for Standards and Controls

Refer to Table 4 for recommended diluents based on different sample types.

In order to meet the lot-specific control ranges provided on the product data
sheet, both the standards and controls should be reconstituted in Bio-Plex

®

standard diluent HB. If reconstituting in a different diluent, users will need to
establish/validate their own control ranges or acceptance criteria.

Table 4. Summary of recommended diluents for standards and controls.

Diluent for Standards

Sample Type

and Controls*

Add BSA

Serum and plasma

Standard diluent HB

None

Culture media, with serum

Culture medium

None

Culture media, serum-free

Culture medium

To 0.5% final**

Lavage, sputum, other fluids

Sample diluent HB

To 0.5% final**

Lysate

Sample diluent HB

To 0.5% final**

* If using diluents other than standard diluent HB, users must establish their own control ranges.
** At least 0.5% final BSA is recommended to stabilize analytes and reduce absorption to labware.