Tables 9–10 summarize the volumes required to prepare 1x detection
antibodies from a single 20x stock. Also shown are volumes to prepare 1x
antibodies when mixing two 20x stocks from singleplex assays.

Preparing 1x detection antibodies from 20x stock
Table 9. Premixed panel or one singleplex assay.

Table 10. Mixing singleplex assays.

5. After incubating the beads, samples, standards, blank, and controls,

slowly remove and discard the sealing tape.

6. Wash the wells three times with 100 µl wash buffer according to the

wash method of choice.

7. Vortex the diluted (1x) detection antibodies gently for 5 sec. Pour into

a reagent reservoir and transfer 25 µl to each well of the assay plate
using a multichannel pipet.

8. Cover and incubate in the dark for 30 min at RT with shaking at

850 ±50 rpm.

20x Detection

Detection Antibody

# of Wells

Antibodies, µl

Diluent, µl

Total Volume, µl

96

145

2,755

2,900

48

73

1,377

1,450

20x Detection

20x Detection

Detection

Antibodies, µl

Antibodies, µl

Antibody

# of Wells

Singleplex #1

Singleplex #2

Diluent, µl

Total Volume, µl

96

145 145 2,610 2,900

48

73 73 1,304 1,450