3 continuous output (stream) format – Rice Lake Intrinsically Safe User Manual

Page 54

Advertising
background image

48

320IS Installation Manual

For example, if the annunciator status value returned on the ZZ command is 148, the gross, standstill, and lb

annunciators are lit: 148 represents the sum of the values for the standstill annunciator (16), gross mode

annunciator (128), and the lb/primary units annunciator (4).

7.3

Continuous Output (Stream) Format

Figure 7-1 shows the continuous output format sent to the

320IS

EDP or printer port when the STREAM parameter

(SERIAL menu) is set to either EDP or PRN.

Figure 7-1. Continuous Output Data Format

Decimal

Value

Annunciator

1

Tare

2

Count

4

lb

8

kg

16

Standstill

32

Center of zero

64

Net

128

Gross

Table 7-3. Status Codes Returned on the ZZ Command

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII

02 (decimal)

Polarity: <Space>
= Positive
<–> = Negative
<^> = Overload
<]> = Underrange

Weight data: 7 digits, right-justified, with
decimal point, leading zero suppression.
Overload =

^^^^^^^

Underrange =

] ] ] ] ] ] ]

Display overflow = OVERFL

L = pounds
K = kilogram
T = ton
G = grams
O = ounces
<space> = none

G = Gross
N = Net

Status: <space> =
valid
I = Invalid
M = In motion
O = Over/under range

<CR> <LF> or

<CR>

ASCII 13, 10

(decimal)