User Manual - Calibration

33

4.0

Calibration

The

320IS

can be calibrated using the front panel, EDP commands, or the Revolution

®

configuration utility. Each

method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional re-zero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

1. Place the indicator in setup mode (display reads

CONFIG

).

2. Remove all weight from the scale platform.
3. If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
4. Press until the display reads

CALIBR

. Press .

WZERO

displays.

5. Press to calibrate zero. The indicator displays

*CAL*

while calibration is in progress. When complete, the

A/D count for the zero calibration is displayed.

DO NOT adjust this value after WSPAN has been set!

6. Press again to save the zero calibration value and go to the next prompt (

WVAL

).

7. Place test weights on the scale.
8. Press to show the test weight value.
9. Use the numeric keypad to enter the actual test weight.
10. Press to save the value.

WSPAN

displays.

11. Press to calibrate span. The indicator displays

*CAL*

while calibrating. When complete, the A/D count

for the span calibration is displayed.

12. Press again to save the span calibration value.

REZERO

displays.

13. Use the rezero function to remove a calibration offset when hooks or chains are used to hang the test

weights.

• If no other apparatus was used to hang the test weights during calibration, remove the test weights and

press to return to the CALIBR menu.

• If hooks or chains were used during calibration, remove these and the test weights from the scale. With all

weight removed, press to rezero the scale. This function adjusts the zero and span calibration values.

The indicator displays

*CAL*

while the zero and span calibrations are adjusted. When complete, the

adjusted A/D count for the zero calibration is displayed. Press to save the value, then press to return to

the CALIBR menu.

14. Press until the display reads

CONFIG

.

15. Press to exit setup mode.

When calibrating to a new scale, after exiting setup mode, press the ZERO key to set Zero and adjust for
deadload.

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit
span calibration
A/D count value

WSPAN

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

VERS

ALGOUT

DIG IN

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

XXXXXXX

SETPNT

Note

Note