Error

Cause

Repair

Out of Range of Impedance

When the subject’s body impedance is
deviated from the limit
- Insufficient touch to electrodes
- Impedance is out of range
- Range: 100 ~ 950 Ω

Clean the measuring parts (the electrodes,
palms, and soles) and try again.

Measure again with correct posture.

Do not move during measurement.

If the same error is repeated, please contact
Rice Lake Weighing Systems (RLWS) or its
local distributor where this device was
purchased.

Out of Range of Body Fat

When the subject’s P.B.F. is deviated from the
limit
- Incorrect input of personal data
- P.B.F. is out of range
- Range: 3 ~ 75 %

Clean the handle electrodes and try again.

If the same error is repeated, please contact
RLWS or its local distributor where this device
was purchased.

Out of Range Measurement

When the subject’s fatness is
deviated from the limit
- Range: -100 ~ +100 %
- Mechanical error

Input height correctly.

Check the weight is correct and try again.

If the same error is repeated, please contact
RLWS or its local distributor where this device
was purchased.

Problem is detected in the
internal communication port.

There’s a problem in the internal
communication port.

Turn off the power and turn on.

If the problem remains, please contact RLWS
or its local distributor where thedevice was
purchased.

Can’t measure the weight

When the subject’s weight is deviated from the
limit
- Measuring error
- Moving during the measurement
- Range: 22.0 ~ 440.9 lb

Measure weight again. If the subject's weight
is out of range, weight can't be measured.

No printing paper

There is no thermal paper.

Insert the thermal paper.

Printer cover is opened

Printer cover is opened.

Check the cover is firmly closed.

Problem is detected in the
Auto-cut of the printer.

Auto-cut blade is shown outward.

Open the cover of Printer-Cut.

Turn the plastic Phillips-head screws clockwise
and push the blade back.

If the problem remains, please contact RLWS
or its local distributor where the device was
purchased.

Problem is detected in the
printer.

Thermal printer has some problems.

Power is automatically turned off by safety
unit.

Turn the power after few minutes. If the
problem remains, please contact RLWS or its
local distributor where the device was
purchased.

Table 7-1. Troubleshooting