17

5-3 Percent Mode (Pct)

This is the mode to display the weight value in percentage compared to a 100% reference mass and

is used for target weighing or checking the sample variable.

Selecting the percent mode

1 Press the MODE key to select pct (percent mode). If the

percent mode can not be selected, refer to “4.
WEIGHING UNITS”.

Storing the 100% reference mass

2 Press the SAMPLE key to enter the 100% reference

mass storing mode.

Even in the storing mode, pressing the MODE key will

switch to the next mode.

3 Place a container on the weighing pan, if necessary.

Press the RE-ZERO key to cancel the weight (tare). The
balance displays 100 0 pct .

4 Place the sample to be set as the 100% reference mass

on the pan or in the container.

5 Press the PRINT key to store the reference mass. The

balance displays 100.00 pct . (The decimal point position
depends on the reference value. The reference mass
stored, even if the AC adapter is removed, is maintained
in non-volatile memory.)

Note

If the balance judges that the mass of the sample is too
light to be used as a reference, it displays
lo . Do not
use the sample.

6 Remove the sample.

Reading the percentage

7 Place a sample to be compared to the reference mass on

the pan. The displayed percentage is based on the 100%
reference mass.

5-4 Animal Weighing Mode (Hold Function)

This is the mode to weigh a moving object such as an animal, even when the display of the weighing
data fluctuates. The hold function allows the average weight of the animal to be displayed.

To use the hold function, set the function in the function table. For details, refer to “9. FUNCTION

TABLE” and “9-4 Description of the Class “Environment, Display””.

Advertising