6 correcting the internal mass value – Rice Lake GX Series, A&D Weighing User Manual

Page 30

Advertising
background image

28

7-6 Correcting the Internal Mass Value

The GX series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.

The internal mass value is corrected as follows:

Model

Correction reference value

Correction range

GX-200
GX-300

200.000 g

GX-400 400.000

g

GX-600
GX-800

GX1000

500.000 g

-0.020 g to +0.020 g

GX-2000
GX-3000
GX-4000
GX-6100

2000.00 g

-0.20 g to +0.20 g

GX-6000
GX-8000

2000.0 g

-2.0 g to +2.0 g

Example: Using the GX-400




















Advertising