7-6 Correcting the Internal Mass Value
The GX series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model
Correction reference value
Correction range
GX-200
GX-300
200.000 g
GX-400 400.000
g
GX-600
GX-800
GX1000
500.000 g
-0.020 g to +0.020 g
GX-2000
GX-3000
GX-4000
GX-6100
2000.00 g
-0.20 g to +0.20 g
GX-6000
GX-8000
2000.0 g
-2.0 g to +2.0 g
Example: Using the GX-400