8. FUNCTION SWITCH AND INITIALIZATION

8-1 Permit or Inhibit

The balance stores parameters that must not be changed carelessly (e.g. Calibration data for
accurate weighing, Data for adapting to the operating environment, Control data for the RS-232C
interface). There are five switches for the purpose of protecting these parameters. Each switch can
select either “permit” or “inhibit”. The “inhibit” protects parameters against careless operations.

Switches

Operation

1 Press the ON:OFF key to turn off the display.

2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The

balance displays p5 .

3 Press the PRINT key. Then the balance displays the function switches.

4 Set the switches using the following keys.

SAMPLE key

To select the switch to change the parameter.

RE-ZERO key

To change the parameter of the switch selected.

0:To inhibit changes. 1:To permit changes

PRINT key

To store the new parameter and return to the weighing mode.

CAL key

To cancel the operation. ( Clr is displayed.)

To return to the weighing mode, press CAL key once again.